Volume 7 Issue 3
May  2016
Turn off MathJax
Article Contents
Li-na GUO, Qi-wen YANG, Jie YI, Xiu-li XIE, Yao WANG, He WANG, Ying-chun XU. Performance Evaluation of Fluorescence Immunochromatography Kit in Detection of Procalcitonin[J]. Medical Journal of Peking Union Medical College Hospital, 2016, 7(3): 208-211. doi: 10.3969/j.issn.1674-9081.2016.03.010
Citation: Li-na GUO, Qi-wen YANG, Jie YI, Xiu-li XIE, Yao WANG, He WANG, Ying-chun XU. Performance Evaluation of Fluorescence Immunochromatography Kit in Detection of Procalcitonin[J]. Medical Journal of Peking Union Medical College Hospital, 2016, 7(3): 208-211. doi: 10.3969/j.issn.1674-9081.2016.03.010

Performance Evaluation of Fluorescence Immunochromatography Kit in Detection of Procalcitonin

doi: 10.3969/j.issn.1674-9081.2016.03.010
More Information
  • Corresponding author: XU Ying-chun   Tel: 010-69159766, E-mail: xycpumch@139.com
  • Received Date: 2015-02-02
  • Publish Date: 2016-05-30
  •   Objective  To evaluate the performance of fluorescence immunochromatography kit in detecting procalcitonin.  Methods  According to the Clinical and Laboratory Standards Institute(CLSI) EP5-A and EP6-A, the performance of fluorescence immunochromatography kit (TEBSUN) in detecting procalcitonin was evaluated in the aspects of precision, linearity, methodology comparison, relative sensitivity, and relative specificity.  Results  The fluorescence immunochromatography kit showed high precision, the coefficients of variation (CV) of low and high concentration samples were 8.3% and 4.7%, respectively. Result of liner verification test showed good linear correlation within the indicated test range (r=0.9989).In methodology comparison tests, the fluorescence immunochromatography kit had good consistency with VIDAS enzyme-linked immunofluorescence kit (r=0.9770).The relative sensitivity and specificity of the fluorescence immunochromatography kit compared with the enzyme-linked immunofluorescence kit were both higher than 86% at 0.5 ng/ml and 2.0 ng/ml level. The total consistency of the two diagnostic kits was 93.75%.  Conclusion  The fluorescence immunochromatography kit has good performance in precision, linearity, methodology comparison, relative sensitivity, and relative specificity, suggesting that it meets the requirements of clinical tests.
  • loading
  • [1] Lee H.Procalcitonin as a biomarker of infectious diseases[J]. Korean J Intern Med, 2013, 28: 285-291. doi:  10.3904/kjim.2013.28.3.285
    [2] Becker KL, Snider R, Nylon ES.Procalcitonin in sepsis and systemic inflammation: a harmful biomarker and a therapeutic target[J]. Br J Pharmacol, 2010, 159:253-264. doi:  10.1111/j.1476-5381.2009.00433.x
    [3] 降钙素原急诊临床应用专家共识组.降钙素原(PCT)急诊临床应用的专家共识[J].中华急诊医学杂志, 2012, 21:944-951. http://www.wanfangdata.com.cn/details/detail.do?_type=perio&id=jzyx201209005
    [4] CLSI.EP5-A. Evaluation of precision performance of clinical chemistry devices[S].Wayne, PA: CLSI, 1999.
    [5] CLSI.EP6-A. Evaluation of the linearity of quantitative measurement procedures: a statistical approach[S].Wayne, PA: CLSI, 2003.
    [6] Morens DM, Fauci AS. Emerging Infectious Diseases: Threats to Human Health and Global Stability[J]. PLoS Pathog, 2013, 9: e1003467. doi:  10.1371/journal.ppat.1003467
    [7] Schneider HG, Lam QT.Procalcitonin for the clinical laboratory: a review[J]. Pathology, 2007, 39:383-390. http://cn.bing.com/academic/profile?id=ce225f32a1faebfae56b18852730f1d1&encoded=0&v=paper_preview&mkt=zh-cn