Li-an HOU, Ling QIU, Qian DI, Xiu-zhi GUO, Hong-jie XIE, Peng-chang LI, Kai WANG, Xin-qi CHENG, Li LIU. Performance Verification of CardioChek PA Lipid Point-of-care Devices[J]. Medical Journal of Peking Union Medical College Hospital, 2014, 5(3): 283-289. doi: 10.3969/j.issn.1674-9081.2014.03.008
Citation: Li-an HOU, Ling QIU, Qian DI, Xiu-zhi GUO, Hong-jie XIE, Peng-chang LI, Kai WANG, Xin-qi CHENG, Li LIU. Performance Verification of CardioChek PA Lipid Point-of-care Devices[J]. Medical Journal of Peking Union Medical College Hospital, 2014, 5(3): 283-289. doi: 10.3969/j.issn.1674-9081.2014.03.008

Performance Verification of CardioChek PA Lipid Point-of-care Devices

doi: 10.3969/j.issn.1674-9081.2014.03.008
More Information
  • Corresponding author: QIU Ling Tel: 010-69159717, E-mail:lingqiubj@aliyun.com
  • Received Date: 2014-05-04
  • Publish Date: 2014-04-30
  •   Objective  To validate the accuracy of CardioChek PA lipid point-of-care devices in determining total cholesterol (TC), triglyceride (TG), and high-density lipoprotein cholesterol (HDL-C) in whole blood and the comparability with results determined by full-automatic biochemical analyzer.  Methods  We determined the low, medium, and high levels of TG, TC, and HDL-C in whole blood with single reagent lot number 20 times using 3 different CardioChek PA devices to evaluate inter-run and devices' coefficient of variations(CVs), and used single CardioChek PA device with 3 reagent lot number to determine the whole blood 10 times for evaluating inter-run and total CVs. Fifty-four volunteers whose lipid profiles covered up high, medium, and low levels were recruited. Lipids in fasting periphery whole blood and venous serum were collected and determined using 3 different CardioChek PA devices and 4 kinds of full-automatic biochemical analyzers, respectively. Bland-Altman plot was made to analyze the comparability of results from CardioChek PA and the full-automatic biochemical analyzers. Linear regression was analyzed using results of CardioChek PA and full-automatic biochemical analyzers. Bias and percentage bias were determined between CardioChek PA and different automatic biochemical analyzers; meanwhile, whether they satisfied the requirements of medical decision levels was determined.  Results  The total CVs of TC in low, medium, and high levels in 3 different devices were 2.69%, 4.88%, and 3.51%, respectively; for TG, they were 5.51%, 5.27% and 4.96%; and for HDL-C, they were 7.27%, 6.84% and 6.79%. The total CVs of TC, TG, and HDL-C determined with the same device but different reagent lot number were 4.70%, 7.66%, and 8.61%, respectively. Comparison of the results from CardioChek PA devices and the 4 kinds of full-automatic biochemical analyzers showed lowest deviation for TC with -2.21%-2.56%, and for HDL-C with -1.12%-5.57%; the deviation of TG results of BeckmanDxC800 was relatively high with 25.85%, but in other 3 systems were -4.55%-13.34%. Deviation of TC, TG, and HDL-C in different medical decision levels were -3.27%-1.96%, -11.05%-13.06%, and -5.86%-11.56%, respectively, all of which could satisfy the requirements of the National Cholesterol Education Program(NCEP). Compared with reference methods, the biases of TC in medical decision levels were 1.96% and 0.77%; for HDL-C, they were 2.34% and 4.87%.  Conclusion  The accuracy of CardioChek PA lipid point-of-care device can satisfy the clinical requirements, and the device can be used in the screening and monitoring of dyslipidemia.
  • loading
  • [1] 陆再英, 钟南山.内科学[M].7版.北京: 人民卫生出版社, 2010: 267-272.
    [2] 陈文祥, 王抒.血脂测定标准化及有关问题[J].浙江实验医学, 2006, 4:31-33. http://www.cnki.com.cn/Article/CJFDTotal-JYYY200601011.htm
    [3] 中国成人血脂异常防治指南制订联合委员会.中国成人血脂异常防治指南(2007版)[J].中华心血管病杂志, 2007, 5:4.
    [4] 李智.POCT技术在临床应用的现状与问题[J].中华检验医学杂志, 2012, 35:1062-1065. http://www.wanfangdata.com.cn/details/detail.do?_type=perio&id=zhyxjy201212002
    [5] Stein JH, Carlsson CM, Papacke-Benson K, et al. Inaccuracy of lipid measurements with the portable Cholestech LDX analyzer in patients with hypercholesterolemia[J]. Clin Chem, 2002, 48:284-290. doi:  10.1093/clinchem/48.2.284
    [6] Clinical and Laboratory Standards Institute. Method comparison and bias estimation using patient samples. EP9-A2[S].Wayne, PA: CLSI, 2002.
    [7] 夏寿扬.改良Bland-Altman图评价血糖仪结果准确性[J].检验医学与临床, 2009, 6:1738-1739. http://www.wanfangdata.com.cn/details/detail.do?_type=perio&id=jyyxylc200920025
    [8] 丛玉隆. POCT的临床应用与存在的问题[J].中华检验医学杂志, 2007, 30:1325-1328. http://www.wanfangdata.com.cn/details/detail.do?_type=perio&id=zhyxjy200712001
    [9] 陶志华.加强医院内部管理保证POCT检验质量[J].中华检验医学杂志, 2012, 35:1077-1080. http://www.wanfangdata.com.cn/details/detail.do?_type=perio&id=zhyxjy201212005
    [10] 华美媛.浅谈血糖POCT的质量管理[J].医学信息学, 2013, 26:6. http://d.wanfangdata.com.cn/Periodical/yxxxzz201315461
  • 加载中

Catalog

    通讯作者: 陈斌, bchen63@163.com
    • 1. 

      沈阳化工大学材料科学与工程学院 沈阳 110142

    1. 本站搜索
    2. 百度学术搜索
    3. 万方数据库搜索
    4. CNKI搜索

    Figures(3)  / Tables(4)

    Article Metrics

    Article views (156) PDF downloads(28) Cited by()
    Proportional views
    Related

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return