Checkmate 2用于放射治疗加速器输出剂量检测的可行性

于浪 杨波 孙玉亮 孙显松 王欣海 李文博 姜斐 李楠 邱杰 张福泉

于浪, 杨波, 孙玉亮, 孙显松, 王欣海, 李文博, 姜斐, 李楠, 邱杰, 张福泉. Checkmate 2用于放射治疗加速器输出剂量检测的可行性[J]. 协和医学杂志, 2016, 7(1): 33-36. doi: 10.3969/j.issn.1674-9081.2016.01.007
引用本文: 于浪, 杨波, 孙玉亮, 孙显松, 王欣海, 李文博, 姜斐, 李楠, 邱杰, 张福泉. Checkmate 2用于放射治疗加速器输出剂量检测的可行性[J]. 协和医学杂志, 2016, 7(1): 33-36. doi: 10.3969/j.issn.1674-9081.2016.01.007
Lang YU, Bo YANG, Yu-liang SUN, Xian-song SUN, Xin-hai WANG, Wen-bo LI, Fei JIANG, Nan LI, Jie QIU, Fu-quan ZHANG. Feasibility of Checkmate 2 in Radiotherapy Accelerator Output Check[J]. Medical Journal of Peking Union Medical College Hospital, 2016, 7(1): 33-36. doi: 10.3969/j.issn.1674-9081.2016.01.007
Citation: Lang YU, Bo YANG, Yu-liang SUN, Xian-song SUN, Xin-hai WANG, Wen-bo LI, Fei JIANG, Nan LI, Jie QIU, Fu-quan ZHANG. Feasibility of Checkmate 2 in Radiotherapy Accelerator Output Check[J]. Medical Journal of Peking Union Medical College Hospital, 2016, 7(1): 33-36. doi: 10.3969/j.issn.1674-9081.2016.01.007

Checkmate 2用于放射治疗加速器输出剂量检测的可行性

doi: 10.3969/j.issn.1674-9081.2016.01.007
详细信息
    通讯作者:

    邱杰 电话:010-69155481, E-mail:qiujie819@aliyun.com

  • 中图分类号: R811.1

Feasibility of Checkmate 2 in Radiotherapy Accelerator Output Check

More Information
  • 摘要:   目的  探讨Checkmate 2作为日检工具用于放射治疗加速器输出剂量检测的可行性。  方法  按照国际原子能机构(International Atomic Energy Agency, IAEA)第277号报告建议方法刻度Varian 2300CD加速器的两档光子线6、15 MV及五档电子线6、9、12、16、20 MeV, 每日分别使用指型电离室和Checkmate 2测量各档能量100 MU的输出量, 得出各自剂量偏差, 最大允许偏差为±2%。分别实测环境温度气压值和读取Checkmate 2温度气压显示值, 计算温度气压修正因子。连续进行3个月工作日共66次测量, 比较每日各档能量测量值以及温度气压修正准确性。  结果  各档能量的电子线及X射线的Checkmate 2与指型电离室测量结果均满足绝对剂量偏差不超过2%的标准, 且两者比较差异均无统计学意义(P > 0.05)。实测及Checkmate 2测量得出的平均温度气压修正因子分别为1.032±0.007和1.033±0.007, 两者差异无统计学意义(P=0.452);相关性分析显示两者呈正相关(r=0.92, P=0.000)。  结论  Checkmate 2可真实反映加速器剂量输出的大体趋势, 且测量工作耗时短、操作简便, 是直线加速器晨检剂量输出检测的良好工具。
  • 图  1  实测及Checkmate 2测得66 d温度气压修正因子趋势线图

    C_ Kpt:实测温度气压修正因子;M_ Kpt:Checkmate 2测量结果得出温度气压修正因子

    表  1  Checkmate 2和指形电离室测量各档能量剂量输出结果(%, x ±s)

    测量方法 6 MeV 9 MeV 12 MeV 16 MeV 20 MeV 6 MV 15 MV
    指形电离室 100.73±0.37 100.51±0.41 100.93±0.42 100.62±0.38 100.85±0.26 100.39±0.43 100.56±0.23
    Checkmate 2 100.66±0.37 100.39±0.41 100.79±0.65 100.62±0.31 100.80±0.36 100.35±0.34 100.52±0.28
    t 1.710 1.750 1.510 0.164 1.501 1.607 1.387
    P 0.920 0.850 0.136 0.870 0.138 0.546 0.170
    下载: 导出CSV
  • [1] Ozyigit G, Yang T, Chao KS, et al. Intensity-modulated radiation therapy for head and neck cancer[J]. Curr Treat Options Oncol, 2004, 5:3-9. doi:  10.1007/s11864-004-0001-y
    [2] Nasr A, Habash A. Dosimetric analytic comparison of inverse and forward planned IMRT techniques in the treatment of head and neck cancer[J]. J Egypt Natl Canc Inst, 2014, 26: 119-125. doi:  10.1016/j.jnci.2014.03.004
    [3] Lee FK, Yip CW, Cheung FC, et al. Dosimetric difference amongst 3 techniques: TomoTherapy, sliding-window intensity-modulated radiotherapy (IMRT), and RapidArc radiotherapy in the treatment of late-stage nasopharyngeal carcinoma (NPC)[J]. Med Dosim, 2014, 39:44-49. doi:  10.1016/j.meddos.2013.09.004
    [4] Chen Z, Errico F, Nath R. Principles and requirements of external beam dosimetry[J]. Radiat Measurements, 2007, 41:S2-S21. http://www.wanfangdata.com.cn/details/detail.do?_type=perio&id=10.1118/1.2761295
    [5] IAEA Technical Reports Series No 277. Absorbed dose determination in photon and electron beam: an international code of practice[R]. 2nd Ed. Vienna: IAEA, 1997.
    [6] 胡逸民.肿瘤放射物理学[M].北京:原子能出版社, 1999.
    [7] Lim S, Ma SY, Jeung TS, et al. Development of a one-stop beam verification system using electronic portal imaging devices for routine quality assurance[J]. Med Dosim, 2012, 37:296-304. doi:  10.1016/j.meddos.2011.11.003