留言板

尊敬的读者、作者、审稿人, 关于本刊的投稿、审稿、编辑和出版的任何问题, 您可以本页添加留言。我们将尽快给您答复。谢谢您的支持!

姓名
邮箱
手机号码
标题
留言内容
验证码

A超角膜测厚仪和Pentacam系统测量散瞳前后角膜中央厚度的比较

金玉梅 艾凤荣 罗岩 李莹

金玉梅, 艾凤荣, 罗岩, 李莹. A超角膜测厚仪和Pentacam系统测量散瞳前后角膜中央厚度的比较[J]. 协和医学杂志, 2013, 4(4): 429-432. doi: 10.3969/j.issn.1674-9081.2013.04.017
引用本文: 金玉梅, 艾凤荣, 罗岩, 李莹. A超角膜测厚仪和Pentacam系统测量散瞳前后角膜中央厚度的比较[J]. 协和医学杂志, 2013, 4(4): 429-432. doi: 10.3969/j.issn.1674-9081.2013.04.017
Yu-mei JIN, Feng-rong AI, Yan LUO, Ying LI. Comparison of Central Corneal Thickness Before and After Mydriasis Measured with A-ultrasound Pachymetry and Pentacam Scheimpflug System[J]. Medical Journal of Peking Union Medical College Hospital, 2013, 4(4): 429-432. doi: 10.3969/j.issn.1674-9081.2013.04.017
Citation: Yu-mei JIN, Feng-rong AI, Yan LUO, Ying LI. Comparison of Central Corneal Thickness Before and After Mydriasis Measured with A-ultrasound Pachymetry and Pentacam Scheimpflug System[J]. Medical Journal of Peking Union Medical College Hospital, 2013, 4(4): 429-432. doi: 10.3969/j.issn.1674-9081.2013.04.017

A超角膜测厚仪和Pentacam系统测量散瞳前后角膜中央厚度的比较

doi: 10.3969/j.issn.1674-9081.2013.04.017
详细信息
    通讯作者:

    李莹 电话:010-69151668, E-mail:liyingpumch@sohu.com

  • 中图分类号: R772.2

Comparison of Central Corneal Thickness Before and After Mydriasis Measured with A-ultrasound Pachymetry and Pentacam Scheimpflug System

More Information
  • 摘要:   目的  比较应用A超角膜测厚仪和Pentacam三维眼前节分析系统测量的散瞳前后角膜中央厚度的差异。  方法  2011年8月至11月北京协和医院成人屈光不正患者65例(130眼), 均先后应用Pentacam三维眼前节分析系统和A超角膜测厚仪进行散瞳前后角膜中央厚度的测量, 散瞳剂采用复方托吡卡胺滴眼液。  结果  A超角膜测厚仪测量的散瞳前后角膜中央厚度值分别为(536.70±39.60)和(548.31±39.22)μm, 散瞳后的测量值明显高于散瞳前(P=0.000);Pentacam三维眼前节分析系统测量的散瞳前后角膜中央厚度值分别为(539.57±39.73)和(545.99±39.47)μm, 散瞳后的测量值亦明显高于散瞳前(P=0.000)。两种方法散瞳前角膜中央厚度的测量值比较, Pentacam三维眼前节分析系统测量值明显高于A超角膜测厚仪(P=0.000);而两种方法散瞳后的测量值比较, 差异无统计学意义(P=0.081)。  结论  散瞳后角膜中央厚度测量值较散瞳前略有增加, 故散瞳前测量值更能准确、真实地反映角膜中央厚度, 有助于为角膜屈光手术方式的选择和手术安全性评价提供参考。
  • [1] Wolfs RC, Klaver CC, Vingerling JR, et al. Distribution of central corneal thickness and its association with intraocular pressure:The Rotterdam Study[J]. Am J Ophthalmol, 1997, 123:767-772. doi:  10.1016/S0002-9394(14)71125-0
    [2] Realini T, Lovelace K. Measuring central corneal thickness with ultrasound pachymetry[J]. Optom Vis Sci, 2003, 80:437-439. doi:  10.1097/00006324-200306000-00009
    [3] Salouti R, Nowroozzadeh MH, Zamani M, et al. Comparison of anterior chamber depth measurements using Calilei, HR Pentacam, and OrbscanⅡ[J]. Optometry, 2010, 81:35-39. doi:  10.1016/j.optm.2009.04.100
    [4] Piňero DP, Alió JL, Alesón A, et al. Corneal volume, pachymetry, and correlation of anterior and posterior corneal shape in subclinical and different stages of clinical keratoconus[J].J Cataract Refract Surg, 2010, 36:814-825. doi:  10.1016/j.jcrs.2009.11.012
    [5] Buehl W, Stojanac D, Sacu S, et al. Comparison of three methods of measuring corneal thickness and anterior chamber depth[J]. Am J Ophthalmol, 2006, 141:7-12. doi:  10.1016/j.ajo.2005.08.048
    [6] Atchison DA. Comparison of peripheral refractions determined by different instruments[J]. Optom Vis Sci, 2003, 80:655-660. doi:  10.1097/00006324-200309000-00011
    [7] Bourges JL, Alfonsi N, Laliberté JF, et al. Average 3-dimensional models for the comparison of Orbscan Ⅱ and Pentacam pachymetry maps in normal corneas[J]. Ophthalmology, 2009, 116:2064-2071. doi:  10.1016/j.ophtha.2009.04.036
    [8] Lattinore MR Jr, Kaupp S, Schallhorn S, et al. Orbscan pachymetry:implications of a repeated measures and diurnal variation analysis[J]. Ophthalmology, 1999, 106:977-981. doi:  10.1016/S0161-6420(99)00519-9
    [9] 叶娅, 沈政伟, 尹禾, 等. Pentacam系统测量正常人眼表面麻醉前后中央角膜厚度的对比研究[J].眼科新进展, 2011, 1:68-70. http://www.wanfangdata.com.cn/details/detail.do?_type=perio&id=ykxjz201101020t
    [10] Rosa N, Lanza M, Borrelli M, et al. Comparison of central corneal thickness measured with Orbscan and Pentacam[J]. J Refract Surg, 2007, 23:895-899. doi:  10.3928/1081-597X-20071101-05
    [11] De Sanctis U, Missolungi A, Mutani B, et al. Reproducibility and repeatability of central corneal thickness measurement in keratoconus using the rotating Scheimpflug camera and ultrasound pachymetry[J]. Am J Ophthalmol, 2007, 144:712-718. doi:  10.1016/j.ajo.2007.07.021
    [12] Ciolino JB, Khachikian SS, Belin MW. Comparison of corneal thickness measurements by ultrasound and Scheimpflug photography in eyes that have undergone laser in situ kerato- mileusis[J]. Am J Ophthalmol, 2008, 145:75-80. doi:  10.1016/j.ajo.2007.08.026
  • 加载中
计量
  • 文章访问数:  95
  • HTML全文浏览量:  40
  • PDF下载量:  0
  • 被引次数: 0
出版历程
  • 收稿日期:  2012-10-30
  • 刊出日期:  2013-10-30

目录

    /

    返回文章
    返回

    【温馨提醒】近日,《协和医学杂志》编辑部接到作者反映,有多名不法人员冒充期刊编辑发送见刊通知,鼓动作者添加微信,从而骗取版面费的行为。特提醒您,本刊与作者联系的方式均为邮件通知或电话,稿件进度通知邮箱为:mjpumch@126.com,编辑部电话为:010-69154261,请提高警惕,谨防上当受骗!如有任何疑问,请致电编辑部核实。谢谢!