应用二维电离室矩阵对后装剂量分布的验证

于浪 杨波 庞廷田 李文博 刘建纯 邱杰 张福泉

于浪, 杨波, 庞廷田, 李文博, 刘建纯, 邱杰, 张福泉. 应用二维电离室矩阵对后装剂量分布的验证[J]. 协和医学杂志, 2013, 4(4): 408-412. doi: 10.3969/j.issn.1674-9081.2013.04.013
引用本文: 于浪, 杨波, 庞廷田, 李文博, 刘建纯, 邱杰, 张福泉. 应用二维电离室矩阵对后装剂量分布的验证[J]. 协和医学杂志, 2013, 4(4): 408-412. doi: 10.3969/j.issn.1674-9081.2013.04.013
Lang YU, Bo YANG, Ting-tian PANG, Wen-bo LI, Jian-chun LIU, Jie QIU, Fu-quan ZHANG. Use of Two-dimensional Ionization Chamber Array on Brachytherapy Dose Verification[J]. Medical Journal of Peking Union Medical College Hospital, 2013, 4(4): 408-412. doi: 10.3969/j.issn.1674-9081.2013.04.013
Citation: Lang YU, Bo YANG, Ting-tian PANG, Wen-bo LI, Jian-chun LIU, Jie QIU, Fu-quan ZHANG. Use of Two-dimensional Ionization Chamber Array on Brachytherapy Dose Verification[J]. Medical Journal of Peking Union Medical College Hospital, 2013, 4(4): 408-412. doi: 10.3969/j.issn.1674-9081.2013.04.013

应用二维电离室矩阵对后装剂量分布的验证

doi: 10.3969/j.issn.1674-9081.2013.04.013
详细信息
    通讯作者:

    邱杰 电话:010-69155481, E-mail:qiujie819@aliyun.com

  • 中图分类号: R811.1

Use of Two-dimensional Ionization Chamber Array on Brachytherapy Dose Verification

More Information
  • 摘要:   目的  探讨二维电离室矩阵系统MatriXX用于近距离治疗剂量分布验证的可行性。  方法  利用井式电离室测量有效测量点60 s的照射量, 结合电离室本身的刻度因子及环境修正因子计算出放射源192Ir的活度。使用该放射源通过Valencia皮肤施源器、等效固体水模, 完成对EBT3免洗胶片的刻度, 获取剂量灰度曲线, 找到经过Valencia皮肤施源器均整后的192Ir辐射野中轴的百分深度剂量(percentage depth dose, PDD)曲线后, 即可对MatriXX进行刻度, 得到剂量刻度因子Kuser。采用Freiburg Flap施源器和插植针及固体水建立模体, 获取CT图像, 在计划系统内完成管道重建以及剂量分布计算, 传输至后装治疗机执行计划, 通过MatriXX获取特定冠状层面内的剂量分布与计划系统所得剂量分布进行比较, 利用软件分析测量与计划剂量分布结果。  结果  经Valencia皮肤施源器均整后的192Ir辐射场的剂量分布类似加速器的射野, 横向及纵向对称性分别为0.17%和0.16%, PDD参考剂量点位于施源器表面下3 mm处, 施源器表面下11.6 mm处的PDD为46.5%。经该辐射场刻度的MatriXX测量的剂量分布与治疗计划系统模体中计算平面相比较, γ分析(3 mm/3%)结果显示, γ ≤ 1的剂量点所占比例为97.67%, 绝对点剂量与治疗计划系统内剂量偏差保持在±5%以内。  结论  MatriXX作为近距离治疗剂量分布的验证工具具有重复性好、避免使用胶片验证的成本损耗等优点, 是目前较为快速、便捷的近距离治疗剂量学质量保证工具。
  • 图  1  Valencia皮肤施源器剖面示意图

    图  2  插入插植针的Freiburg Flap施源器中心与MatriXX中心重合

    图  3  Oncentra计划系统管道重建(A)和剂量分布计算(B)

    图  4  胶片分析所得剂量灰度曲线(A)、射野内对称性(B)及百分深度剂量(C)

    图  5  测量剂量分布(A)及计划系统剂量分布(B)的云图及等剂量线图

    图  6  两个方向轴向的剂量曲线(红色剂量曲线取自测量层面)及γ分析结果A.X方向剂量曲线;B.Y方向剂量曲线;C.γ分析图;D.γ分析结果

  • [1] Haie-Meder C, Pötter R, Van Limbergen E, et al. Recommendations from Gynaecological (GYN) GEC-ESTRO Working Group (Ⅰ):concepts and terms in 3D image-based 3D treatment planning in cervix cancer brachytherapy with emphasis on MRI assessment of GTV and CTV[J]. Radiother Oncol, 2005, 74:235-245. doi:  10.1016/j.radonc.2004.12.015
    [2] Nath R, Anderson LL, Luxton G, et al. Dosimetry of interstitial brachytherapy sources:Recommendations of the AAPM Radiation Therapy Committee Task Group[J]. Med Phys, 1995, 22:209-234. doi:  10.1118/1.597458
    [3] Niroomand-Rad A, Blackwell CR, Coursey BM, et al. Radiochromic film dosimetry:Recommendations of AAPM Radiation Therapy Committee Task Group 55[J].Med Phys, 1998, 25:2093-2115. doi:  10.1118/1.598407
    [4] Herzen J, Todorovic M, Cremers F, et al. Dosimetric evaluation of a 2D pixelionization chamber for implementation in clinical routine[J]. Phys Med Biol, 2007, 52:1197-1208. doi:  10.1088/0031-9155/52/4/023
    [5] Saminathan S, Manickan R, Chandraraj V, et al. Dosimetric study of 2D ion chamber array matrix for the modern radiotherapy treatment verification[J]. J Appl Clin Med Phys, 2010, 11:116-127. doi:  10.1120/jacmp.v11i2.3076
    [6] Wolfsberger LD, Wagar M, Nitsch P, et al. Angular dose dependency of MatriXXTM and its calibration[J]. J Appl Clin Med Phys, 2010, 11:241-251. doi:  10.1120/jacmp.v11i1.3057
    [7] Dobler B, Streck N, Klein E, et al. Hybrid plan verification for intensity-modulated radiation therapy (IMRT) using the 2D ionization chamber array I'mRT MatriXX-a feasibility study[J]. Phys Med Biol, 2010, 55:N39-N55. doi:  10.1088/0031-9155/55/2/N02
    [8] Rickey DW, Sasaki D, Bews J.A quality assurance tool for high-dose-rate braehytherapy[J].Med Phys, 2010, 37:2525-2532. doi:  10.1118/1.3425786
    [9] Poppe B, Djouguela A, Blechschmidt A, et al. Spatial resolution of 2D ionization chamber arrays for IMRT dose verification:single-detector size and sampling step width[J]. Phys Med Biol, 2007, 52:2921-2935. doi:  10.1088/0031-9155/52/10/019
    [10] Poppe B, Mehran P, Kollhoff R, et al. Use of a two-dimensional ionisation chamber array for quality assurance at medical linear accelerators (in German)[J]. Z Med Phys, 2003, 13:115-122. doi:  10.1078/0939-3889-00151
    [11] Granero D, Pérez-Calatayud J, Jimeno J, et al. Design and evaluation of a HDR skin applicator with flattening filter[J]. Med Phys, 2008, 35:495-503. doi:  10.1118/1.2825622
    [12] Devic S, Tomic N, DeBlois F. Commissioning of brachytherapy TPS using EBT model GafChromic film[J]. Med Phys, 2009, 36:4323. doi:  10.1118/1.3244193/abstract
    [13] Sorriaux J, Kacperek A, Rossomme S, et al. Evaluation of Gafchromic EBT3 films characteristics in therapy photon, electron and proton beams[J/OL]. Phys Med, 2012, http://dx.doi.org/10.1016/j.ejmp.2012.10.001.
    [14] Yewondwossen M.Characterization and use of a 2D-array of ion chambers for brachytherapy dosimetric quality assurance[J]. Med Dosim, 2012, 37:250-256. doi:  10.1016/j.meddos.2011.09.004
    [15] Nath R, Anderson LL, Meli JA, et al. Code of practice for brachytherapy physics:report of the AAPM Radiation Therapy Committee Task Group No. 56. American Association of Physicists in Medicine[J]. Med Phys, 1997, 24:1557-1598. doi:  10.1118/1.597966
    [16] Venselaar J, Pêrez-Calatayud J, eds. A practical guide to quality control of brachytherapy equipment[M]. Brussels: ESTRO, 2004.
  • 加载中
图(6)
计量
  • 文章访问数:  176
  • HTML全文浏览量:  47
  • PDF下载量:  5
  • 被引次数: 0
出版历程
  • 收稿日期:  2013-08-21
  • 刊出日期:  2013-10-30

目录

    /

    返回文章
    返回

    【温馨提醒】近日,《协和医学杂志》编辑部接到作者反映,有多名不法人员冒充期刊编辑发送见刊通知,鼓动作者添加微信,从而骗取版面费的行为。特提醒您,本刊与作者联系的方式均为邮件通知或电话,稿件进度通知邮箱为:mjpumch@126.com,编辑部电话为:010-69154261,请提高警惕,谨防上当受骗!如有任何疑问,请致电编辑部核实。谢谢!