骨科手术机器人故障分析及应对策略

刘勋 芮碧宇 张维军 郁红漪 贾英杰

刘勋, 芮碧宇, 张维军, 郁红漪, 贾英杰. 骨科手术机器人故障分析及应对策略[J]. 协和医学杂志, 2023, 14(6): 1149-1154. doi: 10.12290/xhyxzz.2023-0367
引用本文: 刘勋, 芮碧宇, 张维军, 郁红漪, 贾英杰. 骨科手术机器人故障分析及应对策略[J]. 协和医学杂志, 2023, 14(6): 1149-1154. doi: 10.12290/xhyxzz.2023-0367
LIU Xun, RUI Biyu, ZHANG Weijun, YU Hongyi, JIA Yingjie. Analysis and Mitigation Strategies for Failure in Orthopedic Surgical Robots[J]. Medical Journal of Peking Union Medical College Hospital, 2023, 14(6): 1149-1154. doi: 10.12290/xhyxzz.2023-0367
Citation: LIU Xun, RUI Biyu, ZHANG Weijun, YU Hongyi, JIA Yingjie. Analysis and Mitigation Strategies for Failure in Orthopedic Surgical Robots[J]. Medical Journal of Peking Union Medical College Hospital, 2023, 14(6): 1149-1154. doi: 10.12290/xhyxzz.2023-0367

骨科手术机器人故障分析及应对策略

doi: 10.12290/xhyxzz.2023-0367
基金项目: 

国家重点研发计划 2022YFC2409602

国家重点研发计划 2022YFB4700703

详细信息
    通讯作者:

    刘勋, E-mail:liuxun_hust@126.com

  • 中图分类号: R68; TP3

Analysis and Mitigation Strategies for Failure in Orthopedic Surgical Robots

Funds: 

National Key Research and Development Program of China 2022YFC2409602

National Key Research and Development Program of China 2022YFB4700703

More Information
  • 摘要: 伴随人工智能技术取得的巨大突破,骨科手术机器人这一新型医疗器械得以快速发展并具有广阔的应用前景。新技术应用于临床前,拥有极高的可靠性对于保障患者安全和手术质量至关重要,故在研发和临床应用阶段均需对器械的故障类型及其风险有深入认识。本文从研发企业和临床使用者角度出发,全面阐述了骨科手术机器人的故障类型及其发生风险,并针对高风险故障提出了应对策略,旨在为骨科手术机器人器械研发及其临床应用提供指导,助力骨科手术机器人领域高质量发展。
    作者贡献:刘勋负责撰写论文;芮碧宇负责骨科手术机器人临床应用端的故障风险数据收集和整理;张维军和贾英杰负责骨科手术机器人研发端故障风险数据收集和整理;郁红漪负责修订论文。
    利益冲突:所有作者均声明不存在利益冲突
  • 图  1  骨科手术机器人的核心部件

    A.机械臂; B.光学相机; C.跟踪器; D.末端工具; E.电源控制系统; F.工作站与软件系统

    表  1  骨科手术机器人故障类型及设备失效原因

    部件 故障代号 故障类型 设备失效原因
    机械臂 1 系统导航定位精度超差 机械臂定位精度超差
    2 系统导航定位精度超差 机械臂刚度不足,在外力作用下产生变形
    3 系统导航定位错误 机械臂台车未固定,产生意外移动
    4 机械臂碰撞操作者 机械臂产生非预期运动
    5 机械臂碰撞操作者 操作者误入机械臂运动路径
    光学相机 6 系统导航定位精度超差 光学相机精度超差
    7 系统导航定位精度超差 光学相机信号抖动
    跟踪器 8 系统导航定位错误 患者跟踪器发生非预期偏移
    9 系统导航定位错误 手术过程中患者意外移动
    10 中断手术流程 跟踪器被遮挡或超出光学相机检测范围
    末端工具 11 系统导航定位精度超差 末端工具与机械臂连接精度超差
    12 系统导航定位精度超差 末端动力工具超过规划安全边界
    13 医源性感染 末端工具消毒灭菌不彻底
    电源控制系统 14 手术流程中断 输入电源线意外断开
    15 手术流程中断 电源控制系统故障
    工作站与软件系统 16 手术流程中断 软件故障
    17 手术流程中断 通信连接故障
    18 手术流程中断 医学图像质量不能满足手术要求
    下载: 导出CSV

    表  2  研发工程师对骨科手术机器人故障风险等级评价结果前5项

    故障风险等级排序 部件 故障类型 设备失效原因
    1 跟踪器 系统导航定位错误 患者跟踪器发生非预期偏移
    2 末端工具 系统导航定位精度超差 末端工具与机械臂连接精度超差
    3 光学相机 系统导航定位精度超差 光学相机精度超差
    4 末端工具 系统导航定位精度超差 末端动力工具超过规划安全边界
    5 工作站与软件系统 手术流程中断 软件故障
    下载: 导出CSV

    表  3  临床医师对骨科手术机器人故障风险等级评价结果前5项

    故障风险等级排序 部件 故障类型 设备失效原因
    1 跟踪器 系统导航定位错误 患者跟踪器发生非预期偏移
    2 跟踪器 系统导航定位错误 手术过程中患者意外移动
    3 光学相机 系统导航定位精度超差 光学相机精度超差
    4 末端工具 医源性感染 末端工具消毒灭菌不彻底
    5 末端工具 系统导航定位精度超差 末端工具与机械臂连接精度超差
    下载: 导出CSV
  • [1] Kwoh YS, Hou J, Jonckheere EA, et al. A robot with improved absolute positioning accuracy for CT guided stereo-tactic brain surgery[J]. IEEE Trans Biomed Eng, 1988, 35: 153-160. doi:  10.1109/10.1354
    [2] Pransky J. ROBODOC-surgical robot success story[J]. Ind Robot, 1997, 24: 231-233. doi:  10.1108/01439919710167444
    [3] Schulz AP, Seide K, Queitsch C, et al. Results of total hip replacement using the Robodoc surgical assistant system: clinical outcome and evaluation of complications for 97 procedures[J]. Int J Med Robot Comp, 2007, 3: 301-306. doi:  10.1002/rcs.161
    [4] Mazoochian F, Pellengahr C, Huber A, et al. Low accuracy of stem implantation in THR using the CASPAR-system anteversion measurements in 10 hips[J]. Acta Orthop, 2004, 75: 261-264. doi:  10.1080/00016470410001178
    [5] Gomes P. Surgical robotics: Reviewing the past, analysing the present, imagining the future[J]. Robot Cim Int Manuf, 2011, 27: 261-266. doi:  10.1016/j.rcim.2010.06.009
    [6] 朱嘉伟, 杨雪鹤, 郑晨, 等. 中国有源医疗器械行业可靠性及其审评现状研究[J]. 生物医学工程学杂志, 2022, 39: 1050-1058. https://www.cnki.com.cn/Article/CJFDTOTAL-SWGC202205020.htm
    [7] Buchs NC, Pugin F, Volonté F, et al. Reliability of robotic system during general surgical procedures in a university hospital[J]. Am J Surg, 2014, 207: 84-88. doi:  10.1016/j.amjsurg.2013.06.007
    [8] Ren Y, Cao S, Wu J, et al. Efficacy and reliability of active robotic-assisted total knee arthroplasty compared with conventional total knee arthroplasty: a systematic review and meta-analysis[J]. Postgrad Med J, 2019, 95: 125-133. doi:  10.1136/postgradmedj-2018-136190
    [9] 黄健宇. 论FMEA的分析与应用[J]. 机械工业标准化与质量, 2007(4): 37-41.
    [10] 屈媛媛, 王会如, 张宏, 等. 基于FMEA的医疗器械现场检验风险分析[J]. 医疗卫生装备, 2021, 42: 75-79. https://www.cnki.com.cn/Article/CJFDTOTAL-YNWS202101015.htm
    [11] 姚萍华, 王新连. 肺癌根治术体位安置管理中引入失效模式与效应分析的效果[J]. 临床与病理杂志, 2022, 42: 1423-1429.
    [12] Zhang Y, Liu T. Risk assessment based on a STPA-FMEA method: A case study of a sweeping robot[J]. Risk Anal, 2023, 43: 590-604. doi:  10.1111/risa.13927
    [13] Kim KO, Zuo MJ. General model for the risk priority number in failure mode and effects analysis[J]. Reliab Eng Syst Safe, 2018, 169: 321-329. doi:  10.1016/j.ress.2017.09.010
    [14] Zammori F, Gabbrielli R. ANP/RPN: A multi criteria evaluation of the risk priority number[J]. Qual Reliab Eng Int, 2012, 28: 85-104.
  • 加载中
图(1) / 表(3)
计量
  • 文章访问数:  249
  • HTML全文浏览量:  35
  • PDF下载量:  56
  • 被引次数: 0
出版历程
  • 收稿日期:  2023-08-09
  • 录用日期:  2023-11-14
  • 刊出日期:  2023-11-30

目录

    /

    返回文章
    返回

    【温馨提醒】近日,《协和医学杂志》编辑部接到作者反映,有多名不法人员冒充期刊编辑发送见刊通知,鼓动作者添加微信,从而骗取版面费的行为。特提醒您,本刊与作者联系的方式均为邮件通知或电话,稿件进度通知邮箱为:mjpumch@126.com,编辑部电话为:010-69154261,请提高警惕,谨防上当受骗!如有任何疑问,请致电编辑部核实。谢谢!